  • 电感耦合等离子体质谱仪8900.jpg

Inductively Coupled Plasma Mass Spectrometer

Category:

Joint Analysis, Testing, and Technology Cooperation Service Platform

Testing instrument

Process Chemical Testing Instruments

Product accessories:


Contact Us

Product Description

Instrument and Equipment Information

Instrument Name

Inductively Coupled Plasma Mass Spectrometer

Instrument model

8900

Inductively Coupled Plasma Mass Spectrometer

Place of origin and manufacturer

Agilent

Affiliated Unit

Zhenjiang Runjing High-Purity Chemical Technology Co., Ltd.

Acquisition time

Not available yet

Placement location

No. 8 Haixi Road, Zhenjiang New Area

Instrument Status

Normal

Device application range / Instrument measurement range (tolerance)

1. Sensitivity: For low-mass nuclides: Li (7) ≥ 150 Mcps/ppm; for medium-mass nuclides: Y (89) ≥ 500 Mcps/ppm; for high-mass nuclides: Tl (205) ≥ 300 Mcps/ppm or U (238) ≥ 450 Mcps/ppm.

2. Detection limits: Be(9) ≤ 0.1 ppt, In(115) ≤ 0.05 ppt, U(238): 0.05 ppt;

3. Oxide (CeO+/Ce+): ≤1.5%;

4. The ratio of the doubly charged particle (Ce²⁺/Ce) ≤ 3.0%;

5. Abundance sensitivity: Low-mass end: ≤1×10 -9 (i.e., <0.001 ppm); high-quality end: ≤1×10 -9 (i.e., <0.001 ppm);

6. Short-term stability (RSD): <2% (2 min);

7. Long-term stability (RSD): <3% (2 hours);

8. Isotope ratio accuracy: <0.2% ( 107Ag/109 Ag);

9. Anti-interference capability: The detection limit for S (SO+) is <0.05 ppb, and the detection limit for P (PO+) is <0.05 ppb;

10. Background noise: Be9 < 0.3 cps.

Whether it is regularly calibrated by a third-party agency

Yes

CNAS certification obtained.

No