Inductively Coupled Plasma Mass Spectrometer
Category:
Joint Analysis, Testing, and Technology Cooperation Service Platform
Testing instrument
Process Chemical Testing Instruments
Product accessories:
Contact Us
Product Description
|
Instrument and Equipment Information |
|||
|
Instrument Name |
Inductively Coupled Plasma Mass Spectrometer |
Instrument model |
8900 |
|
Inductively Coupled Plasma Mass Spectrometer |
Place of origin and manufacturer |
Agilent |
|
|
Affiliated Unit |
Zhenjiang Runjing High-Purity Chemical Technology Co., Ltd. |
Acquisition time |
Not available yet |
|
Placement location |
No. 8 Haixi Road, Zhenjiang New Area |
Instrument Status |
Normal |
|
Device application range / Instrument measurement range (tolerance) |
1. Sensitivity: For low-mass nuclides: Li (7) ≥ 150 Mcps/ppm; for medium-mass nuclides: Y (89) ≥ 500 Mcps/ppm; for high-mass nuclides: Tl (205) ≥ 300 Mcps/ppm or U (238) ≥ 450 Mcps/ppm. 2. Detection limits: Be(9) ≤ 0.1 ppt, In(115) ≤ 0.05 ppt, U(238): 0.05 ppt; 3. Oxide (CeO+/Ce+): ≤1.5%; 4. The ratio of the doubly charged particle (Ce²⁺/Ce) ≤ 3.0%; 5. Abundance sensitivity: Low-mass end: ≤1×10 -9 (i.e., <0.001 ppm); high-quality end: ≤1×10 -9 (i.e., <0.001 ppm); 6. Short-term stability (RSD): <2% (2 min); 7. Long-term stability (RSD): <3% (2 hours); 8. Isotope ratio accuracy: <0.2% ( 107Ag/109 Ag); 9. Anti-interference capability: The detection limit for S (SO+) is <0.05 ppb, and the detection limit for P (PO+) is <0.05 ppb; 10. Background noise: Be9 < 0.3 cps. |
||
|
Whether it is regularly calibrated by a third-party agency |
Yes |
CNAS certification obtained. |
No |